Analytical Method Development and Validation of Azelnipidine and Telmisartan by RP HPLC Method

V. Spandana*, Siddartha

Department of Pharmaceutical Analysis, Malla Reddy College of Pharmacy, Maisammaguda Dhulapally, Hyderabad, Telangana-500014, India

*Corresponding Author:
V. Spandana
Department of Pharmaceutical Analysis,
Malla Reddy College of Pharmacy,
Maisammaguda, Dhulapally,
Hyderabad,
Telangana-500014,
India,
E-mail:
spandanareddy.vangallu@gmail.com

Received date: 19/11/2021; Accepted date: 03/12/2021; Published date: 10/12/2021

Visit for more related articles at Research & Reviews: Journal of Pharmaceutics and Nanotechnology

Abstract

A simple, Accurate, precise method was developed for the simultaneous estimation of the Azelnipidine and Telmisartan in tablet dosage form. Chromatogram was run through standard symmetry C18 (4.6 × 150 mm, 5 mm). Mobile phase containing Buffer 0.01 N KH2PO4: Acetonitrile taken in the ratio 45:5 % v/v was pumped through column at a flow rate of 1 ml/min. Optimized wavelength selected was 290 nm. Retention time of Azelnipidine and Telmisartan were found to be 2.131 min and 2.593 min. % RSD of the Azelnipidine and Telmisartan were and found to be 0.8 and 0.6 respectively. % Recovery was obtained as 99.59 % and 99.94 % for Azelnipidine and Telmisartan respectively. LOD, LOQ values obtained from regression equations of Azelnipidine and Telmisartan were 0.86, 2.60 and 0.09, 0.29 respectively. Regression equation of Azelnidipine is Y =12533X+10387 and Y=9061X+183.8 of Telmisartan. Retention times were decreased and that run time was decreased, so the method developed was simple and economical that can be adopted in regular Quality control test in Industries.

Keywords

Azelnidipine; Telmisartan ; RP-HPLC

Introduction

Azelnidipine is a compound has a place with the class of natural mixes known as amino benzene sulfonamides. These are natural mixes containing a benzene sulfonamide moiety with an amine gathering joined to the benzene ring (Figure 1).

pharmaceutics-nanotechnology-azelnidipine

Figure 1: Chemical structure of azelnidipine.

Telmisartan, sold under the exchange name Norvir, is an antiretroviral prescription utilized alongside other medications to treat HIV/AIDS. This blend treatment is known as very dynamic antiretroviral treatment (HAART). Regularly a low portion is utilized with other protease inhibitors (Figure 2).

pharmaceutics-nanotechnology-telmisartan

Figure 2: Chemical structure of telmisartan.

The present study was designed to develop a simple, precise, and rapid analytical RP-HPLC procedure, which can be used for the analytical method was developed for the simultaneous estimation of Azelnidipine and Telmisartan as there was only individual methods reported for both drugs. The combination of these two drugs is not official in any pharmacopoeia; hence no official method is available for the simultaneous estimation of these two drugs in their combined dosage forms. Literature survey of Azelnidipine and Telmisartan revealed several methods for detecting these drugs individually but there is only one method for their simultaneous estimation using RP-HPLC [1].

Objective

Following are the objectives of the present work.

• To develop a new stability indicating HPLC method for the simultaneous estimation of Azelnidipine and Telmisartan and to develop the validated method according to ICH guidelines.

• To apply the validated method for the simultaneous estimation of Azelnidipine and Telmisartan in pharmaceutical formulation.

Experimental Methods

Chemicals and reagents

Azelnidipine and Telmisartan pure drugs (API) were from Rankem and combination of Azelnidipine and Telmisartan tablets (durart-r), distilled water, acetonitrile, phosphate buffer, methanol, potassium dihydrogen ortho phosphate buffer and Ortho-phosphoric acid, were purchased from Rankem, Mumbai.

Apparatus and chromatographic condition

Electronics Balance-Denver, PH meter-BVK enterprises, India, Ultrasonicator-BVK enterprises, WATERS HPLC 2695 system equipped with binary pumps, photo diode array detector and Auto sampler integrated with Empower 2 Software. Lab india UV double beam spectrophotometer with UV-win 5 software was used for measuring absorbances of Azelnidipine and Telmisartan The mobile phase was prepared freshly, filtered, sonicated before use and delivered at a flow rate of 1 ml/min and the detector wave length was set at 290 nm injection volume was 1 0 μL. Diluent used was acetonitrile and water taken in the ratio of 50:50 [2].

Preparation of standard and sample solutions standard solution

Accurately weighed 20 mg of Azelnidipine, 2.5 mg of Telmisartan and transferred to 50 ml and 100 ml individual volumetric flask and 3/4th of diluents was added to these flask and sonicated for 10 minutes. Flask were made up with diluents and labeled as Standard stock solution. (400 μg/ml of Azelnidipine and 50 μg/ml of Telmisartan).

Standard working solution

1 ml from each stock solution was pipetted out and taken into a 10 ml volumetric flask and made up with diluent. (40 μg/ml of Azelnidipine and 5 μg/ml of Telmisartan).

Sample stock solution

20 tablets were weighed and calculate the average weight of each tablet then the weight equivalent to one tablet was transferred into a 100 ml volumetric flask. 25 ml of diluent was added, sonicated for 50 min,further the volume made up with diluent and filtered.(4000 μg/ml of Azelnidipine and 500 μg/ml of Telmisartan ). It was centrifuged for 20 min. Then the supernatant was collected and filtered using 0.45 μm filters using (Millipore, Milford, and PVDF). 2 ml from sample stock solution was pipetted out and taken into a 10 ml volumetric flask and made up with diluent. (40 μg/ml of Azelnidipine and 5 μg/ml of Telmisartan).

Procedure

Inject 10 μL of the standard and sample solution separately into the chromatographic system and measure the peak areas for Azelnidipine and Telmisartan and calculate the % assay value [3].

Results and Discussion

All of the analytical validation parameters for this proposed method were determined according to ICH guidelines. Obtained validation parameters are presented in Table 1.

Parameter Azelnidipine Telmisartan
Linearity (μg/ml) 10-60 μg/ml 12.25-7.5 µg/ml
Slope(m) 12533 9061
Intercept(c) 10387 183.8
Regression equation (Y=mX+c) Y=12533X+10387 Y=9061X+183.8
Regression coefficient 0.999 0.999
System precision %RSD 0.7 0.8
Method precision %RSD 0.8 0.6
LOD 0.86 0.09
LOQ 2.60 0.29
Retention Time (min) 2.131 2.593

Table 1. Analytical validation parameters (System suitability and linearity).

Linearity

The calibration curve was constructed by plotting response factor against respective concentration of Azelnidipine and Telmisartan. The plots of peak area vs. respective concentration of Azelnidipine and Telmisartan were found to be linear in the range of 10-60 μg/mL and 1.25-7.5 μg/mL with coefficient of correlation (r2) 0.999 for two drugs. The linearity of this method was evaluated by linear regression analysis. The slope and intercept calculated for Azelnidipine and Telmisartan.

Accuracy

Three levels of Accuracy samples were prepared by standard addition method. Triplicate injections were given for each level of accuracy and mean % Recovery was obtained as 101.14 % and 99.49 % for Azelnidipine and Telmisartan respectively. The obtained results are presented in Table 2.

% Level Amount Spiked (μg/mL) Amount recovered (μg/mL) % RecoverY Mean % Recovery % Level Amount Spiked (μg/mL) Amount recovered (μg/mL) % Recovery Mean %Recovery
  50% 20 19.88 99.42     99.59%   50% 2.5 2.48 99.15     99.94%
20 19.85 99.24 2.5 2.50 99.94
20 19.92 99.62 2.5 2.50 99.92
  100% 40 40.26 100.64   100% 5 5.04 100.86
40 39.84 99.59 5 5.04 100.73
40 39.90 99.75 5 5.04 100.74
  150% 60 59.73 99.54   150% 7.5 7.45 99.40
60 59.50 99.16 7.5 7.45 99.38
60 59.59 99.32 7.5 7.45 99.36

Table 2. Accuracy table of Azelnidipine and Telmisartan.

Sensitivity

The limit of detection (LOD) was determined as lowest concentration giving response and limit of quantification (LOQ) was determined as the lowest concentration analyzed with accuracy of the proposed RP-HPLC method. The limit of detection (LOD) and limit of quantification (LOQ) were found to 0.86 μg/ml and 2.60 μg/ml for Azelnidipine and 0.09 μg/ml and 0.29 μg/ml for Telmisartan. The LOD and LOQ showed that the method is sensitive for Azelnidipine and Telmisartan in Table 3.

Molecule LOD LOQ
Azelnidipine 0.86 2.60
Telmisartan 0.09 0.29

Table 3. Sensitivity table of Azelnidipine and Telmisartan.

System suitability test

The specificity of this method was determined by complete separation of Azelnidipine and Telmisartan as shown in Figure 3 with parameters like retention time, resolution and tailing factor. The tailing factor for peaks of Azelnidipine and Telmisartan was less than 2% and resolution was satisfactory. The average retention time for Azelnidipine and Telmisartan were 2.358 min and 3.099 min respectively for five replicates. The peaks obtained for Azelnidipine and Telmisartan were sharp and have clear baseline separation. Analysis was also performed for active Azelnidipine and Telmisartan, placebo sample (All the ingredients except active Azelnidipine and Telmisartan) both at stressed and unstressed condition. After analysis it was found that there is no interference of peak in the placebo and active sample. Hence the developed method was specific for the analysis of this product [4].

pharmaceutics-nanotechnology-chromatog

Figure 3: Optimized chromatogram of Azelnidipine and Telmisartan.

Precision

From a single volumetric flask of working standard solution six injections were given. A study was carried out for intermediate precision with the same analyst on the different day for six sample preparations of marketed formulations. Robustness of the method was determined by small deliberate changes in flow rate, temperature and mobile phase ratio. The content of the drug was not adversely affected by these changes as evident from the low value of relative standard deviation indicating that the method was rugged and robust. The assay results of tablet dosage formulation by the proposed method are presented in Table 4.

Drug Sample Weight (mg) Inter-day precision System precision Repeatability
SD %
RSD
SD %RSD SD %RSD
Azelnidipine 20 2916.0 0.7 3564.3 0.7 4195.2 0.8
Telmisartan 2.5 2916.0 0.7 285.0 0.7 244.3 0.6

Table 4. Precision of Azelnidipine and Telmisartan.

Stability

In order to demonstrate the stability of both standard and sample solutions during analysis, both solutions were analyzed over a period of 24 hr at room temperature. The results show that for both solutions, the retention time and peak area of Azelnidipine and Telmisartan remained almost similar (% R.S.D. less than 2.0) and no significant degradation within the indicated period, thus indicated that both solutions were stable for at least 24 hr, which was sufficient to complete the whole analytical process. Further forced degradation studies were conducted indicating the stability of the method developed. The results of the degradation studies are presented in Table 5.

  Drug   Label strength (mg)   % Assay
Azelnidipine 400 99.96%
Telmisartan 50 100.45%

Table 5. Assay result of Pharmaceutical dosage formulation.

Assay sample

The contents of Azelnidipine and Telmisartan (40 and 5 mcg each) were collected in 50 ml volumetric flask. Then 20 ml acetonitrile was added, sonicated for 25 min and made up to mark to yield 1110 and 500 μg/ml. It was centrifuged for 20 min. Then the supernatant was collected and filtered using 0.45 μm filters using (Millipore, Milford, PVDF). 2 ml from sample stock solution was pipetted out and taken into a 10 ml volumetric flask and made up with diluent [5]. (40 μg/ml of Azelnidipine and 5 μg/ml of Telmisartan) (Table 6).

Type of degradation Azelnidipine Telmisartan
% Degraded Purity angle Purity threshold % Degraded Purity angle Purity threshold
Acid 3.19 0.21 0.402 4.23 1.267 1.661
Base 4.49 0.11 0.402 2.27 1.662 2.136
Peroxide 2.66 0.21 0.402 2.88 1.501 1.862
Thermal 2 0.126 0.421 4.68 1.244 1.657
Uv 2.28 0.208 0.419 3.27 1.326 1.639
Water 6.27 0.218 0.423 3.24 1.465 1.742

Table 6. Degradation studies of Azelnidipine and Telmisartan.

Acid degradation sample

To 1 ml of stock s solution Azelnidipine and Telmisartan, 1 ml of 2 N Hydrochloric acid was added and refluxed for 30 mins at 60ºc.The resultant solution was diluted to obtain 40 μg/ml and 5 μg/ml solution and 10 μl solutions were injected into the system and the chromatograms were recorded to assess the stability of sample[6]. The typical chromatogram of acid degradation was given in Figure 4.

pharmaceutics-nanotechnology-acid

Figure 4: Optimized chromatogram of azelnidipine and telmisartan acid chromatogram of Azelnidipine and Telmisartan.

Base degradation sample

To 1 ml of stock solution Azelnidipine and Telmisartan, 1 ml of 2 N sodium hydroxide was added and refluxed for 30 mins at 60ºc. The resultant solution was diluted to obtain 40 μg/ml and 5 μg/ml solution and 10 μl were injected into the system and the chromatograms were recorded to assess the stability of sample [7]. The typical chromatogram of acid degradation was given in Figure 5.

pharmaceutics-nanotechnology-working

Figure 5: Base Chromatogram of working sample solution.

Oxidation degradation sample

To 1 ml of stock solution of Azelnidipine and Telmisartan, 1 ml of 20% hydrogen peroxide (H2O2) was added separately. The solutions were kept for 30 min at 60ºc. For HPLC study, the resultant solution was diluted to obtain 40 μg/ml and 5 μg/ml solution and 10 μl were injected into the system and the chromatograms were recorded to assess the stability of sample [8]. The typical chromatogram of oxidative degradation was given in Figure 6.

pharmaceutics-nanotechnology-peroxide

Figure 6: Peroxide chromatogram of Azelnidipine and Telmisartan.

Neutral degradation studies

Stress testing under neutral conditions was studied by refluxing the drug in water for 1 hours at a temperature of 60ºC. For HPLC study, the resultant solution was diluted to 40 μg/ml and 5 μg/ml solution and 10 μl were injected into the system and the chromatograms were recorded to assess the stability of the sample. The typical chromatogram of oxidative degradation was given in Figure 7.

pharmaceutics-nanotechnology-thermal

Figure 7: Thermal chromatogram of Azelnidipine and Telmisartan.

Photo stability studies

The photochemical stability of the drug was also studied by exposing the 250 μg/ml and 625 μg/ml solution to UV Light by keeping the beaker in UV Chamber for 1 day or 200 Watt hours/m2 in photo stability chamber. For HPLC study, the resultant solution was diluted to obtain 40 μg/ml and 5 μg/ml solutions and 10 μl were injected into the system and the chromatograms were recorded to assess the stability of sample. The typical chromatogram of thermal degradation was given in Figure 8.

pharmaceutics-nanotechnology-degradation

Figure 8: UV degradation chromatogram of Azelnidipine and Telmisartan.

Conclusion

A suitable chromatographic method was developed through optimization by changing various parameters such as the mobile phase, injection volume, flow rate etc. In the present method a standard symmetry C18 (4.6 × 150 mm, 5 μm) column has been used For Azelnidipine and Telmisartan respectively. Mobile phase used was Buffer 0.01 N KH2PO4: Acetonitrile (45:55 %v/v) for Azelnidipine and Telmisartan respectively, Retention of Azelnidipine and Telmisartan has more dependence on the mobile phase. The separation of the two peaks was also dependent on the buffer and the percentage of mobile phases. Darunavr and Telmisartan were eluted at acceptable retention times and got good resolution. Several assay methods has been developed for the determination of Azelnidipine and Telmisartan in pharmaceutical dosage forms and in biological fluids but this method is most economic and accurate so this method is very useful for the determination of Azelnidipine and Telmisartan in bulk and pharmaceutical dosage forms. This method was validated as per ICH-Q2 (R1) guidelines and met the regulatory requirements for selectivity, accuracy and stability. Considering the obtained data, it was possible to affirm that the proposed method was fast, simple and suitable for the accurate determination of Azelnidipine and Telmisartan.